Development of a criterion for repositioning a relocatable stereotactic frame using depth helmet measurements
Autor: | Ingrid E.C. van Es, Susan Feith, Esther Raaijmakers |
---|---|
Rok vydání: | 2001 |
Předmět: |
Male
business.industry Computer science Brain Neoplasms Frame (networking) Reproducibility of Results Hematology Middle Aged Radiosurgery Confidence interval Statistics Nonparametric Normal distribution Oncology Humans Radiology Nuclear Medicine and imaging Computer vision Development (differential geometry) Female Head Protective Devices Artificial intelligence business Aged |
Zdroj: | Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 61(1) |
ISSN: | 0167-8140 |
Popis: | Based on 95% confidence intervals in the normal distribution, a criterion was developed for a relocatable stereotactic frame depth helmet measurement. The method for obtaining such a criterion is described. The criterion states that, if more than two points deviate more than 1.5 mm the frame should be repositioned. |
Databáze: | OpenAIRE |