Apical Pressure Generated Using Conventional Syringe Irrigation in Immature Teeth—An In Vitro Study

Autor: Roland Weiger, Ashraf ElAyouti, Marco Jäggi, Florin Eggmann, Thomas Connert, Eva Magni
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: Materials
Materials, Vol 14, Iss 2580, p 2580 (2021)
Volume 14
Issue 10
ISSN: 1996-1944
Popis: This in vitro study aimed to evaluate apical pressure during irrigant delivery with syringe irrigation in immature teeth with an open apical foramen. Conventional syringe irrigation was performed in a 3D-printed immature incisor. A 5 mL syringe combined with 25 G and 30 G cannulas was used. Open-ended and side-vented needle tip designs were assessed. Cannulas were placed at tooth length (TL), TL −1 mm, TL −2 mm, and TL −4 mm. The syringe plunger was moved with a force of 10 N, 20 N, 40 N, and 80 N to simulate clinical conditions. A pressure sensor measured periapical pressures during irrigation. Each experiment was repeated 10 times. Data were analyzed descriptively (maximum, mean, standard deviation, 95% CI) with the critical threshold indicative of extrusion set at 7.64 mbar. 30 G cannulas with both needle tip designs never exceeded the threshold at any TL with a plunger force of 10–40 N. At 80 N, 30 G open-ended cannulas exceeded the threshold in 10%, 30 G side-vented in 20–60% of the measurements. At any TL, 25 G open-ended cannulas and 25 G side-vented cannulas never crossed the threshold with forces of 10–20 N and 10 N, respectively. Consequently, 30 G cannulas with both designs can be recommended for irrigant delivery in immature teeth. 25 G cannulas ought to be used with caution.