Parametric Investigation of an Injection-Jet Self-Powered Fontan Circulation
Autor: | William M. DeCampli, Alain J. Kassab, Arka Das, Ray Prather, Michael Farias, Eduardo Divo |
---|---|
Rok vydání: | 2021 |
Předmět: |
Heart Defects
Congenital Physics Jet (fluid) Multidisciplinary Science Heart Ventricles Hemodynamics Models Cardiovascular Vena Cava Inferior Mechanics Fontan Procedure Fontan circulation Oxygen Child Preschool cardiovascular system Medicine Humans Ventricular Function Computer Simulation Cardiac Output Venous Pressure Parametric statistics |
Zdroj: | Scientific Reports, Vol 12, Iss 1, Pp 1-15 (2022) |
DOI: | 10.21203/rs.3.rs-825168/v1 |
Popis: | Approximately $$1/2500$$ 1 / 2500 babies are born with only one functioning ventricle and the Fontan is the third and, ideally final staged palliative operation for these patients. This altered circulation is prone to failure with survival rates below $$50\%$$ 50 % into adulthood. Chronically elevated inferior vena cava (IVC) pressure is implicated as one cause of the mortality and morbidity in this population. An injection jet shunt (IJS) drawing blood-flow directly from the aortic arch to significantly lower IVC pressure is proposed. A computer-generated 3D model of a 2–4 year old patient with a fenestrated Fontan and a cardiac output of 2.3 L/min was generated. The detailed 3D pulsatile hemodynamics are resolved in a zero-dimensional lumped parameter network tightly-coupled to a 3D computational fluid dynamics model accounting for non-Newtonian blood rheology and resolving turbulence using large eddy simulation. IVC pressure and systemic oxygen saturation were tracked for various IJS-assisted Fontan configurations, altering design parameters such as shunt and fenestration diameters and locations. A baseline “failing” Fontan with a 4 mm fenestration was tuned to have an elevated IVC pressure (+ 17.8 mmHg). Enlargement of the fenestration to 8 mm resulted in a 3 mmHg IVC pressure drop but an unacceptable reduction in systemic oxygen saturation below 80%. Addition of an IJS with a 2 mm nozzle and minor volume load to the ventricle improved the IVC pressure drop to 3.2 mmHg while increasing systemic oxygen saturation above 80%. The salutary effects of the IJS to effectively lower IVC pressure while retaining acceptable levels of oxygen saturation are successfully demonstrated. |
Databáze: | OpenAIRE |
Externí odkaz: |