Halo ring supporting the Brown-Roberts-Wells stereotactic frame for fractionated radiotherapy
Autor: | L. Andreussi, S. Carini, Marcello Ravegnani, Scielzo G, F Bistolfi, F. Grillo Ruggieri |
---|---|
Rok vydání: | 1994 |
Předmět: |
Intracranial Arteriovenous Malformations
Fractionated radiotherapy business.industry Brain Neoplasms Isocenter Equipment Design Fixture Ring (chemistry) Radiosurgery Linear particle accelerator Stereotactic radiotherapy Stereotaxic Techniques Medicine Humans Surgery Neurology (clinical) Halo business Nuclear medicine |
Zdroj: | Acta neurochirurgica. 129(1-2) |
ISSN: | 0001-6268 |
Popis: | The authors describe a new instrumentation for repositioning of the Brown-Roberts-Wells (BRW) stereotaxic system, useful for precise fractionated radiotherapy. A lucite ring is fixed to the patient's skull with four screws. Another ring, partially open, is then firmly connected co-axially to the lower part of the first one with four spacer-bars. The fixture permits an exact repositioning of the B.R.W. stereotaxic system, placing the target point in the linear accelerator isocenter. The preliminary technical results obtained in five children are reported and the fixture performance, advantages, and perspectives are discussed. |
Databáze: | OpenAIRE |
Externí odkaz: |