Image Distortions on a Plastic Interstitial Computed Tomography/Magnetic Resonance Brachytherapy Applicator at 3 Tesla Magnetic Resonance Imaging and Their Dosimetric Impact

Autor: Laura E. van Heerden, Arjan Bel, Kees Koedooder, Zdenko van Kesteren, Oliver J. Gurney-Champion, Antonetta C. Houweling, Coen R. N. Rasch, Bradley R. Pieters
Přispěvatelé: CCA - Imaging and biomarkers, Radiotherapy, Graduate School, Radiology and Nuclear Medicine, CCA -Cancer Center Amsterdam
Rok vydání: 2017
Předmět:
Zdroj: International journal of radiation oncology, biology, physics, 99(3), 710-718. Elsevier Inc.
International Journal of Radiation Oncology-Biology-Physics, 99(3), 710-718. ELSEVIER SCIENCE INC
ISSN: 1879-355X
0360-3016
Popis: Purpose: To quantify magnetic resonance imaging (MRI)distortions on a plastic intracavitary/interstitial applicator with plastic needles at a field strength of 3 T and to determine the dosimetric impact, using patient data.Methods and Materials: For 11 cervical cancer patients, our clinical MRI protocol was extended with 3 scans. From the first scan, a multi-echo acquisition, a map of the magnetic field (B-0) was calculated and used to quantify the field inhomogeneity. The expected displacements of the applicator were quantified for the clinical sequence using the measured field inhomogeneity and the clinical sequence's bandwidth. The second and third scan were our routine clinical sequence (duration
Databáze: OpenAIRE