Pulsatile Pressure Delivery of Continuous-Flow Left Ventricular Assist Devices Is Markedly Reduced Relative to Heart Failure Patients

Autor: Andrew N, Rosenbaum, Timothy L, Rossman, Yogesh N, Reddy, Mauricio A, Villavicencio, John M, Stulak, Philip J, Spencer, Sudhir S, Kushwaha, Atta, Behfar
Rok vydání: 2022
Předmět:
Zdroj: ASAIO Journal. 69:445-450
ISSN: 1058-2916
DOI: 10.1097/mat.0000000000001859
Popis: Although continuous-flow left ventricular assist devices (CF-LVADs) provide an augmentation in systemic perfusion, there is a scarcity of in vivo data regarding systemic pulsatility on support. Patients supported on CF-LVAD therapy (n = 71) who underwent combined left/right catheterization ramp study were included. Aortic pulsatility was defined by the pulsatile power index (PPI), which was also calculated in a cohort of high-output heart failure (HOHF, n = 66) and standard HF cohort (n = 44). PPI was drastically lower in CF-LVAD-supported patients with median PPI of 0.006 (interquartile range [IQR], 0.002-0.012) compared with PPI in the HF population at 0.09 (IQR, 0.06-0.17) or HOHF population at 0.25 (IQR, 0.13-0.37; p0.0001 among groups). With speed augmentation during ramp, PPI values fell quickly in patients with higher PPI at baseline. PPI correlated poorly with left ventricular ejection fraction (LVEF) in all groups. In CF-LVAD patients, there was a stronger correlation with LV dP/dt (r = 0.41; p = 0.001) than LVEF (r = 0.21; p = 0.08; pint0.001). CF-LVAD support is associated with a dramatic reduction in arterial pulsatility as measured by PPI relative to HOHF and HF cohorts and decreases with speed. Further work is needed to determine the applicability to the next generation of device therapy.