Ipsilateral and Simultaneous Comparison of Responses from Acceleromyography- and Electromyography-based Neuromuscular Monitors
Autor: | Edömer Tassonyi, György Nagy, Réka Nemes, Béla Fülesdi, J. Ross Renew, David Robert Hampton, Martyn Gray, Szabolcs Lengyel, Sorin J. Brull |
---|---|
Rok vydání: | 2021 |
Předmět: |
Adult
Male Intraoperative Neurophysiological Monitoring Neuromuscular Junction Pilot Projects Electromyography Clinical study 03 medical and health sciences 0302 clinical medicine Muscle action Tracheal extubation 030202 anesthesiology Accelerometry Medicine Humans Muscle Skeletal medicine.diagnostic_test business.industry Limits of agreement 030208 emergency & critical care medicine Repeatability Middle Aged Neuromuscular monitoring Adductor pollicis muscle Anesthesiology and Pain Medicine Anesthesia Neuromuscular Blockade Female business Neuromuscular Nondepolarizing Agents |
Zdroj: | Anesthesiology. 135(4) |
ISSN: | 1528-1175 |
Popis: | Background The paucity of easy-to-use, reliable objective neuromuscular monitors is an obstacle to universal adoption of routine neuromuscular monitoring. Electromyography (EMG) has been proposed as the optimal neuromuscular monitoring technology since it addresses several acceleromyography limitations. This clinical study compared simultaneous neuromuscular responses recorded from induction of neuromuscular block until recovery using the acceleromyography-based TOF-Watch SX and EMG-based TetraGraph. Methods Fifty consenting patients participated. The acceleromyography and EMG devices analyzed simultaneous contractions (acceleromyography) and muscle action potentials (EMG) from the adductor pollicis muscle by synchronization via fiber optic cable link. Bland–Altman analysis described the agreement between devices during distinct phases of neuromuscular block. The primary endpoint was agreement of acceleromyography- and EMG-derived normalized train-of-four ratios greater than or equal to 80%. Secondary endpoints were agreement in the recovery train-of-four ratio range less than 80% and agreement of baseline train-of-four ratios between the devices. Results Acceleromyography showed normalized train-of-four ratio greater than or equal to 80% earlier than EMG. When acceleromyography showed train-of-four ratio greater than or equal to 80% (n = 2,929), the bias was 1.3 toward acceleromyography (limits of agreement, –14.0 to 16.6). When EMG showed train-of-four ratio greater than or equal to 80% (n = 2,284), the bias was –0.5 toward EMG (–14.7 to 13.6). In the acceleromyography range train-of-four ratio less than 80% (n = 2,802), the bias was 2.1 (–16.1 to 20.2), and in the EMG range train-of-four ratio less than 80% (n = 3,447), it was 2.6 (–14.4 to 19.6). Baseline train-of-four ratios were higher and more variable with acceleromyography than with EMG. Conclusions Bias was lower than in previous studies. Limits of agreement were wider than expected because acceleromyography readings varied more than EMG both at baseline and during recovery. The EMG-based monitor had higher precision and greater repeatability than acceleromyography. This difference between monitors was even greater when EMG data were compared to raw (nonnormalized) acceleromyography measurements. The EMG monitor is a better indicator of adequate recovery from neuromuscular block and readiness for safe tracheal extubation than the acceleromyography monitor. Editor’s Perspective What We Already Know about This Topic What This Article Tells Us That Is New |
Databáze: | OpenAIRE |
Externí odkaz: |