Response of TAPS to monochromatic photons with energies between 45 and 790 MeV

Autor: A.R. Gabler, W. Döring, M. Fuchs, B. Krusche, V. Metag, R. Novotny, M. Röbig-Landau, H. Ströher, V. Tries, C. Molenaar, H. Löhner, J.H.G. van Pol, A. Raschke, M. Šumbera, L.B. Venema, H.W. Wilschut, R. Averbeck, W. Niebur, A. Schubert, R.S. Simon, R. Beck, J. Peise, G.J. Miller, R.O. Owens, G. Anton
Přispěvatelé: Research unit Medical Physics
Rok vydání: 1994
Předmět:
Zdroj: Nuclear instruments & methods in physics research section a-Accelerators spectrometers detectors and associated equipment, 346(1-2), 168-176. ELSEVIER SCIENCE BV
ISSN: 0168-9002
DOI: 10.1016/0168-9002(94)90701-3
Popis: The Two Arm Photon Spectrometer TAPS - comprising 384 plastic-BaF2 scintillator telescopes - was tested with monochromatic photons in the energy range between 45 and 790 MeV. The energy resolution for a collimated photon beam hitting the central detector module was determined to sigma/E = 0.59% X E(gamma)-1/2 + 1.9% (E, given in GeV). For the the fast scintillation component alone sigma/E = 0.79% X E(gamma)-1/2 + 1.8% has been measured. The position resolution of the point of impact amounts to DELTAx = 2 cm (FWHM) at the highest energies which corresponds to 30% of the diameter of an individual module. Monte Carlo simulations using the code GEANT3 are in good agreement with the experimental results.
Databáze: OpenAIRE