Cross-linking to improve THR wear performance

Autor: R.D. Crowninshield, M.P. Laurent, J.Q. Yao, S.K. Bhambri, R.A. Gsell, L.N. Gilbertson, D.F. Swarts
Rok vydání: 2017
Předmět:
Zdroj: Hip international : the journal of clinical and experimental research on hip pathology and therapy. 12(2)
ISSN: 1724-6067
Popis: Polyethylene wear and associated osteolysis can limit the longevity of total hip replacement. In recent years, many improvements have been made in the consolidation, manufacture, and sterilization of polyethylene acetabular components. These improvements provided reduced polyethylene wear and prolonged usefulness of total hip replacement. Recent advances in extensively cross-linking polyethylene offer the possibility to substantially further reduce wear in total hip replacement. Hip simulator wear testing demonstrates an order of magnitude reduction in wear resulting from cross-linking GUR 1050 polyethylene by exposure to 100 kGy of electron beam radiation followed by annealing to encourage cross-linking and to reduce residual free radicals. Clinical investigation will be required to validate the wear advantage of these materials in vivo. (Hip International 2002; 2: 103-7).
Databáze: OpenAIRE