Approach Angle Affects Accuracy in Robotic Stereoelectroencephalography Lead Placement
Autor: | Divaldo Camara, Jordan C. Iordanou, Saadi Ghatan, Fedor Panov |
---|---|
Rok vydání: | 2018 |
Předmět: |
Stereoelectroencephalography
Neurosurgical Procedures Prosthesis Implantation Stereotaxic Techniques 03 medical and health sciences 0302 clinical medicine Robotic Surgical Procedures Medicine Humans Epilepsy surgery Computer vision business.industry Skew Optimal control Magnetic Resonance Imaging Electrodes Implanted Robotic systems Surgery Computer-Assisted 030220 oncology & carcinogenesis Trajectory Robot Surgery Neurology (clinical) Artificial intelligence Electrocorticography business Lead Placement Tomography X-Ray Computed 030217 neurology & neurosurgery |
Zdroj: | World neurosurgery. 128 |
ISSN: | 1878-8769 |
Popis: | One third of patients with epilepsy fail to gain optimal control using antiepileptic drugs. New advances in epilepsy surgery have reshaped some diagnostic and therapeutic modalities into less invasive approaches. To understand the cortical epileptogenic networks, stereoelectroencephalography uses depth electrodes as a tool for invasive intracranial monitoring. These electrodes are now being placed using a robot instead of the traditional Talairach stereotactic grid and frame with comparable accuracy and the advantage of shortened surgical time. Whereas accuracy is important for correct cortical sampling and surgical morbidity, the factors that play a role in the deviation of the electrode's trajectory are not yet understood. The aim of this study was to determine if the angle the planned trajectory makes with the skull impacts the final accuracy of placement.Twenty-five patients with 319 stereotactic trajectories were examined. Skew angle and accuracy measurements were computed based on preoperative and postoperative computed tomography scans fused with trajectories planned using ROSA robot software.Trajectories with a planned angle of30° (oblique group) had significantly higher radial error (2.05 mm vs. 1.45 mm, P0.001) than trajectories with planned angles30° (orthogonal group).The ROSA robotic system is a powerful tool that increases the speed and efficiency of stereotactic neurosurgery. Care should be taken when planning trajectories to avoid high skew angles. If a high skew angle is unavoidable, care should be taken to ensure a larger margin of safety. |
Databáze: | OpenAIRE |
Externí odkaz: |