Misplacement of a Pulmonary Artery Catheter

Autor: Dirk Varelmann, Jan N. Hilberath
Rok vydání: 2014
Předmět:
Zdroj: A & A Case Reports. 3:127-129
ISSN: 2325-7237
DOI: 10.1213/xaa.0000000000000079
Popis: Placing a flow-directed pulmonary artery catheter (PAC) can be difficult and lead to serious complications. We present the case of an attempted PAC insertion in a patient undergoing implantation of a left ventricular assist device. Although physiologic pressure waveforms were established, plausible measurements of cardiac output and pulmonary capillary wedge pressure were not initially obtainable. Chest radiography showed that the catheter was looped in the pulmonary artery. This malposition could not be visualized intraoperatively by transesophageal echocardiography. The PAC was replaced in the operating room after the left ventricular assist device was implanted.
Databáze: OpenAIRE