A new applicator design for endocavitary brachytherapy of cancer in the nasopharynx

Autor: Dick Sipkema, Cees A. Meeuwis, Leo L. Visch, Connie de Pan, Rob Peters, Peter C. Levendag, Paul I.M. Schmitz
Přispěvatelé: Radiotherapy, Erasmus MC other
Rok vydání: 1997
Předmět:
Zdroj: Radiotherapy and Oncology, 45, 95-98. Elsevier Ireland Ltd
ISSN: 0167-8140
Popis: Introduction : In attempting to improve local tumor control by higher doses of radiation, there has been a resurgence of interest in the implementation of brachytherapy in the management of primary and recurrent cancers of the nasopharynx. Brachytherapy with its steep dose fall-off is of particular interest because of the proximity of critical dose limiting structures. Recent developments in brachytherapy, such as the introduction of pulsed-dose-rate and high-dose-rate computerized afterloaders, have encouraged further evolution of brachytherapy techniques. Materials and methods : We have designed an inexpensive, re-usable and flexible silicone applicator, tailored to the shape of the soft tissues of the nasopharynx, which can be used with either low-dose-rate brachytherapy or high (pulsed)-dose-rate remote controlled afterloaders. Results and conclusions : This Rotterdam nasopharynx applicator proved to be easy to introduce, patient friendly and can remain in situ for the duration of the treatment (2–6 days). The design, technique of application and the first consecutive 5 years of clinical experience in using this applicator are presented.