Trends in Fidelis Lead Survival

Autor: David J. Slotwiner, Apoor Patel, Aviva Tobin-Hess, Bruce G. Goldner, Jim W. Cheung
Rok vydání: 2012
Předmět:
Zdroj: Circulation: Arrhythmia and Electrophysiology. 5:906-912
ISSN: 1941-3084
1941-3149
DOI: 10.1161/circep.112.972000
Popis: Background— The Sprint Fidelis implantable cardioverter-defibrillator lead was recalled in 2007 because of an elevated risk of lead fracture. Several studies have demonstrated an accelerating risk of lead failure over time. We sought to identify predictors and characterize trends of Fidelis lead failure. Methods and Results— We evaluated 604 Fidelis leads with ≥90 days of follow-up implanted at our institution. Fidelis lead survival was analyzed by the Kaplan-Meier method. Analysis of log-log plots of cumulative hazard plots was performed to assess changes in lead failure rate over time. During follow-up of 3.3±1.7 years, 51 (8.4%) Fidelis lead failures were identified. The 3-year and 5-year Fidelis lead survival rates were 93.5% and 85.3%, respectively. Female sex was the only significant predictor of lead failure (heart rate, 2.1; 95% CI, 1.1–3.9; P P Conclusions— After 3 years, failure rates of Fidelis leads stabilize but at a significantly elevated rate. Female sex is associated with a doubling of the risk of Fidelis lead failure. These findings have implications for Fidelis lead management decisions that are based on the prediction of lead failure risk.
Databáze: OpenAIRE