Identification of diethylene glycol monobutyl ether as a source of contamination in an ion trap mass spectrometer
Autor: | Christopher R. Gibson, Cheri M. Brown |
---|---|
Jazyk: | angličtina |
Předmět: |
Chromatography
Tandem Liquid scintillation counting Ether Contamination Mass spectrometry High-performance liquid chromatography Mass Spectrometry Molecular Weight chemistry.chemical_compound chemistry Pharmaceutical Preparations Structural Biology Ethylene Glycols Indicators and Reagents Ion trap Radiopharmaceuticals Spectroscopy Chromatography High Pressure Liquid Diethylene glycol monobutyl ether |
Zdroj: | Journal of the American Society for Mass Spectrometry. (11):1247-1249 |
ISSN: | 1044-0305 |
DOI: | 10.1016/S1044-0305(03)00534-8 |
Popis: | Tandem liquid chromatography-mass spectrometry coupled to online radioactive material detection (LC/RAM/MS/MS) is a technique that is used routinely for in vivo and in vitro drug metabolism studies and allows for a simultaneous correlation between radiochemical peaks and mass spectral data. The compound diethylene glycol monobutyl ether (DGBE), a component of a commercially available scintillation cocktail for RAM analysis, was identified as a source of overwhelming chemical noise in a mass spectrometer which was used in an LC/RAM/MS/MS configuration. In this report, we describe the identification of DGBE as the source of the chemical noise and the methods that were used to minimize the exposure of the mass spectrometer to volatile components of the scintillation cocktail. |
Databáze: | OpenAIRE |
Externí odkaz: |