A 27μW subcutaneous wireless biosensing platform with optical power and data transfer

Autor: Kannan Aryaperumal Sankaragomathi, Babak A. Parviz, Ramin Mirjalili, Brian Otis, Luis Perez
Rok vydání: 2014
Předmět:
Zdroj: CICC
DOI: 10.1109/cicc.2014.6946089
Popis: We present a batteryless, 27uW barely subcutaneous sensing platform using optical power and data links. We demonstrate an 8.1 mm × 3.2 mm implantable tag powered by an 850nm infrared source through the skin barrier. Measurements using pig skin indicate that optical power transfer through skin achieves a 4.9% efficiency, which is higher than mm scale inductive power links. As a proof-of-concept demo, we sense and transmit temperature to an external body worn reader using a fully optical power and data link.