Langzeitergebnisse nach Implantation einer Plattenhaptik-Silikonlinse in den Kapselsack
Autor: | Martin Zehetmayer, Herbert Weghaupt, Christian Skorpik, Thomas Pfleger, U. Scholz |
---|---|
Rok vydání: | 1994 |
Předmět: |
medicine.medical_specialty
Visual acuity genetic structures business.industry medicine.medical_treatment Intraocular lens Cataract surgery Astigmatism medicine.disease eye diseases Surgery Ophthalmology chemistry.chemical_compound Silicone chemistry medicine Capsulotomy sense organs Implant medicine.symptom business Macular edema |
Zdroj: | Klinische Monatsblätter für Augenheilkunde. 204:220-225 |
ISSN: | 1439-3999 0023-2165 |
DOI: | 10.1055/s-2008-1035521 |
Popis: | BACKGROUND An important step in developing cataract surgery was the introduction of soft foldable silicone intraocular lenses in the middle 80's. Functional and morphological long-term data are a vital base for definite conclusions on the new material's safety and biocompatibility. PATIENTS AND METHODS Our study presents the long-term results after the implantation of a silicone posterior chamber lens with solid plate haptic design (STAAR AA-4203). All lenses were folded and implanted through a 4-mm small corneoscleral incision. Our study includes 54 eyes in 52 patients with a mean follow-up period of 56.5 +/- 8.9 months. RESULTS A visual acuity of 20/40 or better was found in 90.7% of all eyes. All except one of the 33 eyes without any further preexisting ocular pathology at the time of the operation achieved a visual acuity of 20/40 or better. Two thirds of all eyes had a horizontal astigmatism of +1.0 D or less. By slitlamp examination more or less dispersed pigment was seen on the IOL's surface in 46.3%. 83% of the silicone lenses were centrated within 0.5 mm. A YAG-laser capsulotomy was performed in 7 cases (13%). IOL-related intraocular inflammations, cystoid macular edema or an elevated intraocular pressure did not occur in this group. CONCLUSIONS Our satisfactory long-term results suggest an excellent biocompatibility of the IOL's design and material. |
Databáze: | OpenAIRE |
Externí odkaz: |