Development of High-resolution TV Camera for Real Time Digital Radiography System
Autor: | Masayuki Tsuneoka, Fumitaka Takahashi, Yokouchi Hisatake, Yoichi Onodera |
---|---|
Rok vydání: | 1993 |
Předmět: | |
Zdroj: | The Journal of the Institute of Television Engineers of Japan. 47:732-738 |
ISSN: | 1884-9652 0386-6831 |
DOI: | 10.3169/itej1978.47.732 |
Popis: | We have developed a 2000-line TV camera which use a 1-inch SATICON as a real-time digital radiography (DR) systemWe record a 60% amplitude response when 1000 TV lines are at the center of the image, and a 46 dB signal-to-noise ratio using the 2100 line scanning mode at a signal current of 550 nA |
Databáze: | OpenAIRE |