Gamma-Index Passing Rates in Baseline Plans Measured with a Detector Array
Autor: | Kirpal Kohli, Stanislaw Szpala |
---|---|
Rok vydání: | 2015 |
Předmět: | |
Zdroj: | International Journal of Medical Physics, Clinical Engineering and Radiation Oncology. :326-337 |
ISSN: | 2168-5444 2168-5436 |
DOI: | 10.4236/ijmpcero.2015.44039 |
Popis: | Purpose: This study provides a simple protocol for validation of the gamma passing rates and to identify the optimum values of % dose and mm criteria for dose distributions measured with a detector array. Methods: We chose ArcCHECK detector array to illustrate the concepts. We used plans with uniform or quasi-uniform dose distributions along the detector array for testing in the presence of dose errors. For testing sensitivity to spatial shift we employed a plan with approximately constant dose gradient along the axis of the instrument. Results: We identified a representative set of parameters which describe performance of a detector array. We determined the minimum gamma-index acceptance criteria allowing the passing rates to reach 100% in the absence of errors, and identified the minimum fully detectable errors for such criteria. For our baseline plans delivered to ArcCHECK, 100% passing rates were obtained for 1.5% dose criterion together with ±3% minimum error detectable at 100% rate, and for 1.5 mm criterion together with the minimum fully detectable error of ±3 mm. We inspected the impact of selected program options on the passing rates. Conclusions: The protocol we developed provides a simple method of commissioning-style analysis of a detector array without a need for analysis of a large number of clinical plans. |