Autor: |
J. Kyle Polack, Alice Tomanin, Shaun D. Clarke, Paolo Peerani, M.M. Bourne, Sara A. Pozzi, J. L. Dolan, C.C. Lawrence, Marek Flaska |
Rok vydání: |
2014 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 767:188-192 |
ISSN: |
0168-9002 |
DOI: |
10.1016/j.nima.2014.08.002 |
Popis: |
Neutron measurements can be used to distinguish plutonium in metal or oxide form, a capability that is of great interest in nuclear nonproliferation, treaty verification, and other applications. This paper describes measurements performed on well-characterized samples of plutonium oxide and plutonium metal using the pulse-shape-discrimination-capable plastic scintillator EJ-299-33. Results are compared to those obtained with a same-sized detector cell using the liquid scintillator EJ-309. The same optimized, digital pulse shape discrimination technique is applied to both detectors and the neutron pulse height distributions are compared. Results show that the EJ-299-33 plastics can be successfully used for plutonium measurements, where the gamma ray to neutron detection ratio is much higher than for typical radioactive sources. Results also show that EJ-299-33 detectors can be used to characterize plutonium samples, specifically to discriminate between plutonium metal and oxide. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|