X‐ray gauge measures areal density variations as small as 0.1%

Autor: David M. Stupin
Rok vydání: 1985
Předmět:
Zdroj: Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films. 3:1266-1268
ISSN: 1520-8559
0734-2101
DOI: 10.1116/1.573080
Popis: A gauge measures areal density variations as small as 0.1% in low density plastic foams and other materials. It has a spatial resolution of 0.5 mm. This system is safe, portable, easy to use, and is readily automated. The 6 keV x-ray flux from a /sup 55/Fe source is collimated to a 0.5 mm diameter beam and attenuated by the sample. Transmitted x-rays are detected with a proportional gas detector and counted with a scaler-timer. After measuring the x-ray attenuation in two areas of the sample, the areal density variation is calculated using an