Notice of Removal: Design and in-vivo evaluation of next-generation laparoscopic HIFU kidney probe

Autor: Christie Johnson, Ralf Seip, Roy F. Carlson, Jesse Clanton, Adam Morris, Jay Morris, Liang Cheng, Mahesh Matam, Mark Carol, Jennifer A. Smith, Rodrigo Chaluisan, Narendra T. Sanghvi, Sameer Chopra, Inderbir S. Gill, Jacob Carr, Chandru P. Sundaram, Cheuk Fan Shum
Rok vydání: 2017
Předmět:
Zdroj: 2017 IEEE International Ultrasonics Symposium (IUS).
DOI: 10.1109/ultsym.2017.8092872
Popis: Previously, we reported on the development and clinical evaluation of an ultrasound image-guided laparoscopic high-intensity focused ultrasound (HIFU) probe for kidney tumor ablation compatible with an 18mm trocar. Since then, the relentless march towards miniaturization of laparoscopic tools has led to the obsolescence of this trocar size, prompting a re-design of the probe to accommodate market realities and user expectations. Reducing the probe to be compatible with a 15mm trocar proved challenging, as simply scaling the existing design would have resulted in clinically unacceptable ablation performance.