AFRIMETS.AUV.A-S2 final report, calibration of a multi-function acoustic calibrator
Autor: | R Nel, H Potgieter, D Dobrowolska, A Młyńska, H A Shawky, A Maina, K Ojasalo |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | Metrologia. 60:09003 |
ISSN: | 1681-7575 0026-1394 |
Popis: | Main text Results of Regional Metrology Organisation supplementary comparison AFRIMETS.AUV.A-S2 concerning the calibration of a Brüel & Kjær 4226 multi-frequency sound calibrator are reported here. Sound pressure level results at frequencies from 31.5 Hz to 16 kHz and nominal sound level settings of 94 dB and 114 dB, were evaluated for consistency, the calculated comparison reference values and Degrees of Equivalence are reported here. The comparison's purpose was to prove technical capabilities supporting the submissions of calibration measurement capabilities in the BIPM key comparison database. The light shines quite far, the new calibration measurement capabilities that may result from this work underpin some important fundamental acoustic calibration services, e.g., sound calibrators, sound level meters and personal sound exposure meters. These instruments are intrinsic to the effective monitoring and implementation worldwide of human hearing conservation and hearing loss assessment programs, all of which impacts quality of life. According to the World Health Organization, nearly 1.5 billion people experience some decline in their hearing capacity during their lives [1]. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). |
Databáze: | OpenAIRE |
Externí odkaz: |