Rhenium-186 has been standardized for activity by the CIEMAT/NIST method of 4πβ liquid-scintillation efficiency tracing with tritium, with an uncertainty (equivalent to 1 SD) of 1.61%. Half-life measurements with a pressurized ionization chamber give a T 1 2 = 89.25 ± 0.07 h . Photon emission rates were measured with semiconductor detectors, and the probability of emission of the principal γ-ray at 137 keV was found to be P7 = 0.0945 ± 0.0016. |