Use of the Radionics Image Fusion™ and Stereoplan™ programs for target localization in functional neurosurgery
Autor: | Jeremy Rowe, Vakis Papanastassiou, Richard B. Scott, L E Davies, Tipu Z. Aziz, Peter A. Silburn |
---|---|
Rok vydání: | 1998 |
Předmět: |
medicine.medical_specialty
Image fusion medicine.diagnostic_test business.industry Thalamotomy medicine.medical_treatment Magnetic resonance imaging General Medicine Functional neurosurgery Surgery Neurology Radionics Physiology (medical) Medicine Functional surgery Pallidotomy Neurology (clinical) One pass business Nuclear medicine |
Zdroj: | Journal of Clinical Neuroscience. 5:28-32 |
ISSN: | 0967-5868 |
DOI: | 10.1016/s0967-5868(98)90197-7 |
Popis: | We describe the use of Radionics Image Fusion™ and Stereoplan™ in defining the target for thalamotomy and pallidotomy in functional surgery for parkinsonism and tremor. Using this to fuse and spatially correct magnetic resonance imaging (MRI) to computed tomography (CT) images our calculated targets were a mean of 0.6 ± 1.5 mm from the end target determined physiologically by stimulation. This is significantly better than the values of 2.6 ± 1.6 mm for thalamic targets and 7.1 ± 3.7 mm for pallidal targets using CT alone. As a consequence, determination of the target and the lesion making are routinely performed in one pass of the electrode allowing for faster, more accurate and, we believe, safer functional procedures. |
Databáze: | OpenAIRE |
Externí odkaz: |