CO2laser stabilized on narrow saturated absorption resonances of CO2, improved absolute frequency measurements
Autor: | V. Bernard, G. Nogues, L. F. Constantin, Ch. Chardonnet, Ch. Daussy |
---|---|
Rok vydání: | 1997 |
Předmět: | |
Zdroj: | Metrologia. 34:313-318 |
ISSN: | 1681-7575 0026-1394 |
Popis: | We have stabilized a CO2 laser on narrow saturated absorption resonances of CO2 detected on the transmitted beam of an external Fabry-Perot cavity in the 30 THz spectral region. By using another CO2 laser stabilized on a close OsO4 line whose absolute frequency is precisely known, and measuring the beat frequency 4 with a precision of 100 Hz, we determined the absolute frequency of twenty CO2 laser lines, with uncertainties between 100 Hz and 1 kHz mainly limited by the accuracy of the OsO4 grid. The set of measurements reveal some discrepancies with data based on measurements performed with the widely used saturated fluorescence technique; we attribute these to pressure shifts of several kilohertz under the conditions of this last method. |
Databáze: | OpenAIRE |
Externí odkaz: |