CZT detectors with 3D readout for gamma-ray spectroscopy and imaging

Autor: Robert T. Skelton, Michael R. Pelling, James L. Matteson
Rok vydání: 2003
Předmět:
Zdroj: X-Ray and Gamma-Ray Detectors and Applications IV.
ISSN: 0277-786X
DOI: 10.1117/12.455978
Popis: We are developing 10 mm thick CZT detectors with 3-D readout for ~100 keV to ~1.5 MeV gamma-rays. Multiple-site gamma-ray interactions are fully measured, i.e., the energy and 3-D position of each site are determined. Spatial resolution is 1 mm FWHM. Anode pixel readout with 1 mm pitch is used for x- and y-positions and charge drift times for z-positions. Drift time measurements are triggered by the cathode signal and end when each interaction site's charge cloud reaches an anode pixel. Post-event processing corrects for signal loss due to charge trapping and accurately determines gamma-ray energies, with a goal of 1% energy resolution at 662 keV. Compton kinematic analysis can identify the initial interaction site in most cases as well as constrain the incident gamma-ray direction. Tests were made with a prototype detector, measuring 10 x 10 x 10 mm3 and operated at 1000 V bias. The measured drift time resolution of 25 nsec FWHM at 662 keV and 60 nsec at 122 keV corresponds to z-position resolution of 0.25 and 0.60 mm FWHM, respectively. The technique is described and results of modeling and tests are presented.