A method for determination of hydrogen content in uranium dioxide samples using a semiconductor sensor
Autor: | S. D. Lazarev, S. S. Yakimov, N. E. Teterin, V. Ya. Goncharov, N. P. Zvereva, V. I. Filippov |
---|---|
Rok vydání: | 1991 |
Předmět: |
Hydrogen
Health Toxicology and Mutagenesis Semiconductor materials Hydrogen molecule Uranium dioxide Public Health Environmental and Occupational Health Analytical chemistry chemistry.chemical_element Hydrogen content Pollution Analytical Chemistry chemistry.chemical_compound Nuclear Energy and Engineering chemistry Radiology Nuclear Medicine and imaging Gas detector Trace analysis Semiconductor sensor Spectroscopy |
Zdroj: | Journal of Radioanalytical and Nuclear Chemistry Articles. 147:197-200 |
ISSN: | 1588-2780 0236-5731 |
Popis: | Measurements of H2 content in UO2 samples have been carried out on the basis of a technique using a hydrogen sensitive sensor of Pd−SiO2−Si. A specially made set up allowed to register the amount of H2 in UO2 samples up to 10−10, g. Measured H2 content in investigated samples was 8, 32 and 102 ppb (mass) at 400, 800 and 1600°C, respectively. |