336. Validation of a methodology for SPECT/CT absolute quantification
Autor: | Franca Foppiano, O. Ferrando, A. Chimenz, Andrea Ciarmiello |
---|---|
Rok vydání: | 2018 |
Předmět: |
Contouring
Scanner business.industry Absolute quantification Instrumentation Biophysics General Physics and Astronomy General Medicine Imaging phantom 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine 030220 oncology & carcinogenesis Calibration Analysis software Radiology Nuclear Medicine and imaging Nuclear medicine business Correction for attenuation Mathematics |
Zdroj: | Physica Medica. 56:265-266 |
ISSN: | 1120-1797 |
Popis: | Purpose To evaluate the activity quantification performance of our SPECT/CT scanner (SYMBIA T2 – SIEMENS) and to verify a methodology for absolute activity quantification. Methods Quantitative SPECT/CT studies can be performed using the methodology proposed by Zeintl et al. [1] . The instrumentation consists of an uniform cylindrical phantom (5680 mL) and an IEC NEMA torso-phantom with hot sphere inserts. 99mTc was used to prepare the radioactive solutions. Different sphere-to-background ratio (RSB) were used (61:1, 44:1, 33:1, 10:1). Images were reconstructed using the proprietary OSEM 3D algorithm with resolution recovery (FLASH 3D, Siemens Healthcare), CT-based attenuation correction and energy window-based scatter correction. A system calibration factor (S) relating the count rate (cps) and the activity (mCi) contained in a tridimensional volume was determined together with the activity recovery coefficients. The phantom results were extended to 99mTc-MAA patient acquisitions for activity quantification in liver lesions. Lesions contouring was performed using the Volumetric Analysis Software (Siemens Healthcare). Results A calibration factor of 17190 cps/mCi was obtained for 99mTc. Recovery coefficients (Fig. 1) vary from 90% (for volumes ⩾25 mL) to 20% (for volumes up to 0.5 mL). For 128 updates and no post-smoothing filtering we obtain the maximum activity recovery (about 90%) for a volume ⩾25 mL. Hot sphere contrast and activity quantification improve with higher updates while the sphere to background ratio and the background variability increase. A balance between activity recovery and background variability is obtained with 8 iterations, 8subset and no filtering. Applying the phantom results to 7 MAA-99mTc acquisitions, activity accumulated in liver lesions can be determined with an accuracy ⩽10% (Table 1). Conclusions With the proposed methodology spect activity quantification can be reached with an accuracy ⩽10% for 99mTc acquisitions. In the future the method will be tested also with other isotopes used in the clinical practice. |
Databáze: | OpenAIRE |
Externí odkaz: |