Quantitative T2 mapping using accelerated 3D stack-of-spiral gradient echo readout
Autor: | Qin Qin, Ruoxun Zi, Dan Zhu |
---|---|
Rok vydání: | 2020 |
Předmět: |
Data consistency
Computer science Noise (signal processing) Biomedical Engineering Biophysics Partial volume Pulse sequence Iterative reconstruction 030218 nuclear medicine & medical imaging 03 medical and health sciences 0302 clinical medicine Compressed sensing Consistency (statistics) Radiology Nuclear Medicine and imaging Algorithm 030217 neurology & neurosurgery Spiral |
Zdroj: | Magnetic Resonance Imaging. 73:138-147 |
ISSN: | 0730-725X |
DOI: | 10.1016/j.mri.2020.08.007 |
Popis: | Purpose To develop a rapid T2 mapping protocol using optimized spiral acquisition, accelerated reconstruction, and model fitting. Materials and methods A T2-prepared stack-of-spiral gradient echo (GRE) pulse sequence was applied. A model-based approach joined with compressed sensing was compared with the two methods applied separately for accelerated reconstruction and T2 mapping. A 2-parameter-weighted fitting method was compared with 2- or 3-parameter models for accurate T2 estimation under the influences of noise and B1 inhomogeneity. The performance was evaluated using both digital phantoms and healthy volunteers. Mitigating partial voluming with cerebrospinal fluid (CSF) was also tested. Results Simulations demonstrates that the 2-parameter-weighted fitting approach was robust to a large range of B1 scales and SNR levels. With an in-plane acceleration factor of 5, the model-based compressed sensing-incorporated method yielded around 8% normalized errors compared to references. The T2 estimation with and without CSF nulling was consistent with literature values. Conclusion This work demonstrated the feasibility of a T2 quantification technique with 3D high-resolution and whole-brain coverage in 2–3 min. The proposed iterative reconstruction method, which utilized the model consistency, data consistency and spatial sparsity jointly, provided reasonable T2 estimation. The technique also allowed mitigation of CSF partial volume effect. |
Databáze: | OpenAIRE |
Externí odkaz: |