High resolution FROG system for the characterization of ps laser pulses
Autor: | Johannes A. L'huillier, D. Woll, Harry Fuchs, Thorsten Ulm |
---|---|
Rok vydání: | 2007 |
Předmět: |
Materials science
Physics and Astronomy (miscellaneous) Frequency-resolved optical gating Spectrometer business.industry General Engineering General Physics and Astronomy Laser Radius of curvature (optics) law.invention Semiconductor laser theory Optics law Picosecond Spectral width Spectral resolution business |
Zdroj: | Applied Physics B. 88:393-396 |
ISSN: | 1432-0649 0946-2171 |
DOI: | 10.1007/s00340-007-2730-7 |
Popis: | We report on a frequency resolved optical gating (FROG) system which allows the reconstruction of amplitude and phase of picosecond pulses generated by mode-locked diode laser oscillators and amplifiers. The main challenge of developing such a FROG system is the required high spectral resolution. In the spectrometer a cylindrical grating with 1200 lines/mm and a radius of curvature of 3 m is used. A width of the illuminated area of 15 cm provides in a spectral resolution exceeding 2 GHz (1.4 pm) at 460 nm. The FROG system is, thus, well suited for analyzing ps-pulses with durations of up to 80 ps and a corresponding fourier-limited spectral width of 0.017 nm. |
Databáze: | OpenAIRE |