A fast and accurate calibration method for high sensitivity pressure transducers
Autor: | Heng Yang, Yuancheng Sun, Minhang Bao, Jinsuo Wang |
---|---|
Rok vydání: | 2003 |
Předmět: |
Engineering
Calibration curve business.industry Acoustics Metals and Alloys Repeatability Condensed Matter Physics Pressure sensor Surfaces Coatings and Films Electronic Optical and Magnetic Materials law.invention Compensation (engineering) Pressure measurement law Range (statistics) Calibration Electronic engineering Sensitivity (control systems) Electrical and Electronic Engineering business Instrumentation |
Zdroj: | Sensors and Actuators A: Physical. 108:218-223 |
ISSN: | 0924-4247 |
DOI: | 10.1016/s0924-4247(03)00266-8 |
Popis: | A novel calibration scheme for fast accurate calibration of high sensitivity pressure transducers is proposed and verified by experiments. Based on theoretical analysis, the accuracy of the calibration for non-linearity, hysteresis and repeatability can be much higher than 1×10−4 for pressure transducers with an operation range of 1 kPa or below. For accurate calibration of sensitivity, a single conventional measurement is needed at the full range pressure. A conventional compensation pressure gauge can be used for the new calibration method. The calibration is very accurate and very fast; only a few seconds for a test point (or 10–20 min for a device). As the operation pressure is self-generating, no expensive pressure source is needed. The performances can be further improved if the design of the apparatus is optimized for the new calibration method according to the theoretical analysis. |
Databáze: | OpenAIRE |
Externí odkaz: |