Observation of X-ray shadings in synchrotron radiation-total reflection X-ray fluorescence using a color X-ray camera

Autor: Ursula E. A. Fittschen, Velma Lopez, George J. Havrilla, Uwe Reinholz, Oliver Scharf, G. Buzanich, Kathryn G. McIntosh, Martin Radtke, Magnus Menzel, Christina Streli
Rok vydání: 2014
Předmět:
Zdroj: Spectrochimica Acta Part B: Atomic Spectroscopy. 99:179-184
ISSN: 0584-8547
DOI: 10.1016/j.sab.2014.06.025
Popis: Absorption effects and the impact of specimen shape on TXRF analysis has been discussed intensively. Model calculations indicated that ring shaped specimens should give better results in terms of higher counts per mass signals than filled rectangle or circle shaped specimens. One major reason for the difference in signal is shading effects. Full field micro-XRF with a color X-ray camera (CXC) was used to investigate shading, which occurs when working with small angles of excitation as in TXRF. The device allows monitoring the illuminated parts of the sample and the shaded parts at the same time. It is expected that sample material hit first by the primary beam shade material behind it. Using the CXC shading could be directly visualized for the high concentration specimens. In order to compare the experimental results with calculation of the shading effect the generation of controlled specimens is crucial. This was achieved by “drop on demand” technology. It allows generating uniform, microscopic deposits of elements. The experimentally measured shadings match well with those expected from calculation.
Databáze: OpenAIRE