Défaut d'isolement des laryngoscopes de marque « Upsher » série UQ1

Autor: B. Vailly, L. Tritsch
Rok vydání: 2006
Předmět:
Zdroj: Annales Françaises d'Anesthésie et de Réanimation. 25:69-70
ISSN: 0750-7658
DOI: 10.1016/j.annfar.2005.07.083
Popis: We observed an electrical short-circuit between a fasten screw of the printed circuit and the handle of an Upsher universal laryngoscope (serial number UQ1). The isolating Silicone layer was broken above the screw. This isolation defect was found all over our Upsher laryngoscopes of the UQ1 series. No doubt that if accumulators were used instead of batteries, emitted heat would be in largest amount and perhaps dangerous.
Databáze: OpenAIRE