A study of the relation between the limit of detection and the limit of quantitation in inductively coupled plasma spectrochemistry

Autor: M. Carré, S. Excoffier, Jean-Michel Mermet
Rok vydání: 1997
Předmět:
Zdroj: Spectrochimica Acta Part B: Atomic Spectroscopy. 52:2043-2049
ISSN: 0584-8547
DOI: 10.1016/s0584-8547(97)00114-6
Popis: The limit of quantitation based on a repeatability threshold concept is compared with the limit of detection in induction coupled plasma atomic emission spectrometry (ICP-AES) and induction coupled plasma mass spectrometry (ICP-MS). A 5%-based limit of quantitation would normally correspond to 10 times the 3-σ based limit of detection. However, because of a possible lack of linearity of the calibration graph at low concentrations, some additional noise not taken into consideration and the possible use of time-correlated multichannel detection, this ratio of 10 cannot be used in every case. It is suggested that a suitable way of determining the limit of quantitation is to establish the plot of the percentage relative standard deviation (RSD) of the net signal as a function of the concentration in a range from the limit of detection to 50 times this limit.
Databáze: OpenAIRE