SU-E-J-185: Gated CBCT Imaging for Positioning Moving Lung Tumor in Lung SBRT Treatment
Autor: | Dwight E. Heron, D.A. Clump, M.S. Huq, Steven A. Burton, Tianfang Li, Y Zhang, Xiang Li, Brian J Karlovits |
---|---|
Rok vydání: | 2014 |
Předmět: |
Cone beam computed tomography
medicine.medical_specialty Lung business.industry medicine.medical_treatment Image processing General Medicine Iterative reconstruction respiratory system urologic and male genital diseases equipment and supplies Radiation therapy medicine.anatomical_structure stomatognathic system medicine Lung tumor Tomography Radiology Nuclear medicine business Projection (set theory) |
Zdroj: | Medical Physics. 41:199-199 |
ISSN: | 0094-2405 |
Popis: | Purpose: Lung stereo-tactic body radiotherapy(SBRT) treatment requires high accuracy of lung tumor positioning during treatment, which is usually accomplished by free breathing Cone-Beam computerized tomography (CBCT) scan. However, respiratory motion induced image artifacts in free breathing CBCT may degrade such positioning accuracy. The purpose of this study is to investigate the feasibility of gated CBCT imaging for lung SBRT treatment. Methods: Six Lung SBRT patients were selected for this study. The respiratory motion of the tumors ranged from 1.2cm to 3.5cm, and the gating windows for all patients were set between 35% and 65% of the respiratory phases. Each Lung SBRT patient underwent free-breathing CBCT scan using half-fan scan technique. The acquired projection images were transferred out for off-line analyses. An In-house semi-automatic algorithm was developed to trace the diaphragm movement from those projection images to acquire a patient's specific respiratory motion curve, which was used to correlate respiratory phases with each projection image. Afterwards, a filtered back-projection algorithm was utilized to reconstruct the gated CBCT images based on the projection images only within the gating window. Results: Target volumes determined by free breathing CBCT images were 71.9%±72% bigger than the volume shown in gated CBCT image. On the contrary, the target volume differences between gated CBCT and planning CT images at exhale stage were 5.8%±2.4%. The center to center distance of the targets shown in free breathing CBCT and gated CBCT images were 9.2±8.1mm. For one particular case, the superior boundary of the target was shifted 15mm between free breathing CBCT and gated CBCT. Conclusion: Gated CBCT imaging provides better representation of the moving lung tumor with less motion artifacts, and has the potential to improve the positioning accuracy in lung SBRT treatment. |
Databáze: | OpenAIRE |
Externí odkaz: |