Autor: |
Joao Varela, R. Moura, Miguel Ferreira, José C. Da Silva, Pedro Amaral, B. Carrico, Andreia Trindade, Catarina Ortigao, Pedro Rodrigues |
Rok vydání: |
2007 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 580:1123-1126 |
ISSN: |
0168-9002 |
Popis: |
Clear-PEM is a dedicated PET scanner for breast and axilla cancer diagnosis, under development within the framework of the Crystal Clear Collaboration at CERN, aiming at the detection of tumors down to 2 mm in diameter. The camera consists of two planar detector heads with active dimensions 16.0×14.5 cm 2 . Each head has 96 Clear-PEM detector modules consisting of 32 LYSO:Ce pixels with dimensions 2×2×20 mm 3 packed in a 4×8 BaSO 4 reflector matrix compressed between two Hamamatsu S8550 APD arrays in a double-readout configuration for Depth-of-Interaction (DoI) determination. The modules are individually measured and characterized before being grouped into Supermodules (comprised of 24 modules). Measured properties include photo-peak position, relative gain dispersion, energy resolution, cross-talk and DoI resolution. Optical inspection of matrices was also performed with the aid of a microscope, to search for pixel misalignments and matrix defects. Modules’ performance was thoroughly evaluated with a 511 keV collimated beam to exactly determine DoI resolution. In addition, a fast quality control (QC) procedure using flood irradiations from a 137 Cs source was applied systematically. The overall performance of the 24 detector modules complies with the design goals of the Clear-PEM detector, showing energy resolution around 15%, DoI resolution of about 2 mm and gain dispersion among pixels of 15%. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|