Compact x-ray spectrometer based on thermoluminescent detectors
Autor: | I. G. Grigoryeva, M. V. Khil'ko, G. Kh. Salakhutdinov |
---|---|
Rok vydání: | 2019 |
Předmět: | |
Zdroj: | Journal of Physics: Conference Series. 1390:012105 |
ISSN: | 1742-6596 1742-6588 |
DOI: | 10.1088/1742-6596/1390/1/012105 |
Popis: | A small-sized pulsed x-ray spectrometer was designed and developed. The spectrometer is designed to study pulsed x-rays in the energy range of quanta 1-30 keV. The spectrometer is an assembly of thermoluminescent LiF detectors arranged one after another. Thermoluminescent detectors record x-ray radiation and are used as filters for selection of x-ray radiation in the spectrum. Thermoluminescent LiF detectors are immune to electromagnetic interference, and their response is linear in a wide dynamic range of the absorbed radiation dose (from 20 mSv to 10 Sv). Calibration measurements and test tests have shown good performance and reliability of the small-sized spectrometer. |
Databáze: | OpenAIRE |
Externí odkaz: |