Holocamera for 3-D micrography of the alert human eye

Autor: Tokuda, Allan R., Auth, David C., Bruckner, Adam P.
Zdroj: Applied Optics; July 1980, Vol. 19 Issue: 13 p2219-2225, 7p
Abstrakt: A holocamera that safely records holograms of the full depth of the alert human eye with a spatial resolution of ∼20 μm is described. A single-mode argon-ion laser generating 2 W at 5145 Å serves as the illuminating source. Holographic exposure times of 0.3 msec are achieved by means of a fail-safe electromechanical shutter system. Integrated retinal irradiance levels are well under the American National Standards Institute safety standards. Reconstructed real images are projected directly onto the vidicon faceplate of a closed-circuit TV system, enabling convenient scanning in the x-y-z dimensions of the reconstructed eyeball. Serially reconstructed holograms of cataractous rabbit eyes and normal human eyes are presented.
Databáze: Supplemental Index