Ishikiriyama, K., Wunderlich, B. |
Zdroj: |
Journal of Thermal Analysis & Calorimetry; Sep1997, Vol. 50 Issue 3, p337-346, 10p |
Abstrakt: |
The quality of measurement of heat capacity by differential scanning calorimetry (DSC) is based on strict symmetry of the twin calorimeter. This symmetry is of particular importance for temperature-modulated DSC (TMDSC) since positive and negative deviations from symmetry cannot be distinguished in the most popular analysis methods. The heat capacities for sapphire-filled and empty aluminum calorimeters (pans) under designed cell imbalance caused by different pan-masses were measured. In addition, the positive and negative signs of asymmetry have been explored by analyzing the phase-shift between temperature and heat flow for sapphire and empty runs. The phase shifts change by more than 180° depending on the sign of the asymmetry. Once the sign of asymmetry is determined, the asymmetry correction for temperature-modulated DSC can be made. [ABSTRACT FROM AUTHOR] |