Autor: |
Janssens-bij de Vaate, Mieke, Rhebergen, Koenraad S. |
Předmět: |
|
Zdroj: |
International Journal of Audiology; Jul2024, Vol. 63 Issue 7, p527-534, 8p |
Abstrakt: |
In standards IEC 60645-1 and ANSI S3.6, the free-field equivalent earphone output level method is assumed as the reference for speech audiometry. Three calibration procedures for this method were compared in this study. Speech audiometry was conducted with Dutch consonant-vowel-consonant words for the following conditions: 1. TDH39 earphones, 2. loudspeaker, and 3. free-field simulated with TDH39 earphones. The first calibration procedure was based on the empirically determined difference between the speech recognition threshold (SRT) with earphones and a loudspeaker. The second procedure was based on the theoretical free-field correction, derived from the known speech spectrum and the free-field to coupler difference. The third calibration procedure corresponded to the results of the free-field simulated speech material under earphones. The sample included 20 normal hearing subjects. The differences between the observed SRT in the free-field and earphone conditions and the free-field and simulated free-field conditions were 7.1 dB and 0.6 dB, respectively. The three calibration procedures for the free-field equivalent output method yielded approximately the same results, and therefore all appear to be useful for TDH39 earphones. [ABSTRACT FROM AUTHOR] |