Autor: |
Frank, G., Tyers, O., Larrieu, Alberto J., Nishimura, Akira, Williams, Edward H., Kurusz, Mark, Brownlee, Robert R. |
Předmět: |
|
Zdroj: |
Pacing & Clinical Electrophysiology; Jan1980, Vol. 3 Issue 1, p84-87, 4p |
Abstrakt: |
A ventricular inhibited demand cardiac pacemaker was inhibited in the presence of intimate contact between the active temporary bipolar electrode ring and a rejoined inactive permanent bipolar electrode tip. Electromagnetic interference effects, lack of insulation, lead breaks, and loose connections were all ruled out as the cause of pacemaker suppression. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |