Autor: |
Williams, David L., Brown, Craig M., Tong, David, Sulyman, Alexander, Gary, Charles K. |
Předmět: |
|
Zdroj: |
Journal of Imaging; Dec2020, Vol. 6 Issue 12, p1-11, 11p |
Abstrakt: |
Resolution measurements were made using 14.1 MeV neutrons from a high-yield, portable DT neutron generator and a neutron camera based on a scintillation screen viewed by a digital camera. Resolution measurements were made using a custom-built, plastic, USAF-1951 resolution chart, of dimensions 125 × 98 × 25.4 mm3, and by calculating the modulation transfer function from the edge-spread function from edges of plastic and steel objects. A portable neutron generator with a yield of 3 × 109 n/s (DT) and a spot size of 1.5 mm was used to irradiate the object with neutrons for 10 min. The neutron camera, based on a 6LiF/ZnS:Cu-doped polypropylene scintillation screen and digital camera was placed at a distance of 140 cm, and produced an image with a spatial resolution of 0.35 cycles per millimeter. [ABSTRACT FROM AUTHOR] |
Databáze: |
Complementary Index |