Monte Carlo Simulations and Measurements for Efficiency Determination of Lead Shielded Plastic Scintillator Detectors.

Autor: Yasin, Zafar, Negoita, Florin, Tabbassum, Sana, Borcea, Ruxandra, Kisyov, Stanimir
Předmět:
Zdroj: AIP Conference Proceedings; 2017, Vol. 1916 Issue 1, p1-5, 5p, 1 Diagram, 4 Graphs
Abstrakt: The plastic scintillators are used in different areas of science and technology. One of the use of these scintillator detectors is as beam loss monitors (BLM) for new generation of high intensity heavy ion in superconducting linear accelerators. Operated in pulse counting mode with rather high thresholds and shielded by few centimeters of lead in order to cope with radiofrequency noise and X-ray background emitted by accelerator cavities, they preserve high efficiency for high energy gamma ray and neutrons produced in the nuclear reactions of lost beam particles with accelerator components. Efficiency calculation and calibration of detectors is very important before their practical usage. In the present work, the efficiency of plastic scintillator detectors is simulated using FLUKA for different gamma and neutron sources like, 60Co, 137Cs and 238Pu-Be. The sources are placed at different positions around the detector. Calculated values are compared with the measured values and a reasonable agreement is observed. [ABSTRACT FROM AUTHOR]
Databáze: Complementary Index