Dosimetry of ultra-high dose rate electron beams using thermoluminescence and optically stimulated luminescence detectors.
Autor: | Motta S; Department of Radiation Safety and Security, Paul Scherrer Institute, Villigen PSI, Switzerland.; Department of Physics, ETH Zürich, Zürich, Switzerland., Dal Bello R; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland., Christensen JB; Department of Radiation Safety and Security, Paul Scherrer Institute, Villigen PSI, Switzerland., Bossin L; Department of Radiation Safety and Security, Paul Scherrer Institute, Villigen PSI, Switzerland., Yukihara EG; Department of Radiation Safety and Security, Paul Scherrer Institute, Villigen PSI, Switzerland. |
---|---|
Jazyk: | angličtina |
Zdroj: | Physics in medicine and biology [Phys Med Biol] 2024 Feb 02; Vol. 69 (3). Date of Electronic Publication: 2024 Feb 02. |
DOI: | 10.1088/1361-6560/ad1cf5 |
Abstrakt: | Objective. The aim of this work is to investigate the dose rate dependence of thermoluminescence and optically stimulated luminescence detectors (TLDs and OSLDs) in a wide uniform ultra-high dose rate electron beam and demonstrate the potential use of TLDs and OSLDs to correct the ion recombination in air-filled ionization chambers. This study avoids previously reported complications related to the field size and homogeneity. Approach. Two types of OSLDs (BeO and Al (Creative Commons Attribution license.) |
Databáze: | MEDLINE |