Biomechanical design optimization of proximal humerus locked plates: A review.

Autor: Zdero R; Orthopaedic Biomechanics Lab, Victoria Hospital, London, ON, Canada., Brzozowski P; Orthopaedic Biomechanics Lab, Victoria Hospital, London, ON, Canada. Electronic address: pawel.brzozowski@lhsc.on.ca., Schemitsch EH; Orthopaedic Biomechanics Lab, Victoria Hospital, London, ON, Canada; Division of Orthopaedic Surgery, Western University, London, ON, Canada.
Jazyk: angličtina
Zdroj: Injury [Injury] 2024 Feb; Vol. 55 (2), pp. 111247. Date of Electronic Publication: 2023 Nov 28.
DOI: 10.1016/j.injury.2023.111247
Abstrakt: Background: Proximal humerus locked plates (PHLPs) are widely used for fracture surgery. Yet, non-union, malunion, infection, avascular necrosis, screw cut-out (i.e., perforation), fixation failure, and re-operation occur. Most biomechanical investigators compare a specific PHLP configuration to other implants like non-locked plates, nails, wires, and arthroplasties. However, it is unknown whether the PHLP configuration is biomechanically optimal according to some well-known biomechanical criteria. Therefore, this is the first review of the systematic optimization of plate and/or screw design variables for improved PHLP biomechanical performance.
Methods: The PubMed website was searched for papers using the terms "proximal humerus" or "shoulder" plus "biomechanics/biomechanical" plus "locked/locking plates". PHLP papers were included if they were (a) optimization studies that systematically varied plate and screw variables to determine their influence on PHLP's biomechanical performance; (b) focused on plate and screw variables rather than augmentation techniques (i.e., extra implants, bone struts, or cement); (c) published after the year 2000 signaling the commercial availability of locked plate technology; and (d) written in English.
Results: The 41 eligible papers involved experimental testing and/or finite element modeling. Plate variables investigated by these papers were geometry, material, and/or position, while screw variables studied were number, distribution, angle, size, and/or threads. Numerical outcomes given by these papers included stiffness, strength, fracture motion, bone and implant stress, and/or the number of loading cycles to failure. But, no paper fully optimized any plate or screw variable for a PHLP by simultaneously applying four well-established biomechanical criteria: (a) allow controlled fracture motion for early callus generation; (b) reduce bone and implant stress below the material's ultimate stress to prevent failure; (c) maintain sufficient bone-plate interface stress to reduce bone resorption (i.e., stress shielding); and (d) increase the number of loading cycles before failure for a clinically beneficial lifespan (i.e., fatigue life). Finally, this review made suggestions for future work, identified clinical implications, and assessed the quality of the papers reviewed.
Conclusions: Applying biomechanical optimization criteria can assist biomedical engineers in designing or evaluating PHLPs, so orthopaedic surgeons can have superior PHLP constructs for clinical use.
Competing Interests: Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
(Copyright © 2023 Elsevier Ltd. All rights reserved.)