Experimentally-guided development of a sensor for pleural cavity pressure sensing.

Autor: Mimra T, Cerny M, Guerin C, Noury N
Jazyk: angličtina
Zdroj: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2022 Jul; Vol. 2022, pp. 4245-4248.
DOI: 10.1109/EMBC48229.2022.9871731
Abstrakt: The precise monitoring of the intrapleural pressures, necessitates the development of a specific instrumental approach to select the correct shape, dimensions and material to implement the sensing balloon which will be inserted into the pleural cavity. We demonstrate that a 10 cm diameter disk, printed with filaments of TPU (Thermoplastic polyurethane) with hardness 92A offers the best compromise in terms of static sensitivity (0.28 mV/cmHg) and dynamic frequency response (48 Hz).