Two-dimensional shear wave elastography with two different systems for the diagnosis of breast lesions.
Autor: | Yu JF; Department of Medical Ultrasound, Shanghai Tenth Hospital, School of Clinical Medicine of Nanjing Medical University, Shanghai, P.R. China.; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Zhang S; Department of Medical Ultrasound, Shanghai Tenth Hospital, School of Clinical Medicine of Nanjing Medical University, Shanghai, P.R. China.; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Yin HH; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Zhou BG; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Pu YY; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Fang Y; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Du D; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Zhang Y; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China., Xu HX; Department of Medical Ultrasound, Shanghai Tenth Hospital, School of Clinical Medicine of Nanjing Medical University, Shanghai, P.R. China.; Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital; Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University. Shanghai, P.R. China.; Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment; National Clinical Research Center for Interventional Medicine, Shanghai, P.R. China. |
---|---|
Jazyk: | angličtina |
Zdroj: | Clinical hemorheology and microcirculation [Clin Hemorheol Microcirc] 2022; Vol. 82 (1), pp. 53-62. |
DOI: | 10.3233/CH-221471 |
Abstrakt: | Background: Two-dimensional (2D) - shear wave elastography (SWE) has made promising advances in the diagnostic of breast lesions. However, few studies have assessed whether the diagnostic effectiveness of different platforms employing 2D-SWE is equal or different. Objective: To compare the diagnostic effectiveness of 2D-SWE techniques from two different systems in differentiating malignant breast lesions from benign ones. Methods: A total of 84 breast lesions were retrospectively analyzed by experienced radiologists using 2D-SWE on two ultrasound systems, i.e. system-1 (LOGIQ E9 system, GE Healthcare, Wauwatosa, WI, USA), and system-2 (Aixplorer US system, SuperSonic Imagine, Aix-en-Provence, France). Qualitative and quantitative parameters including color sign, the maximum elasticity modulus values (E-max), the mean elasticity modulus values (E-mean) and standard deviation (E-sd) of elasticity modulus values in two 2D-SWE systems were analyzed. The diagnostic performance between system-1 and system-2 were evaluated in terms of the areas under the receiver operating characteristic curves (AUROCs). Results: Among the 84 lesions in this study, 66 (78.6%) were benign and 18 (21.4%) were malignant. E-max in system-1 showed the best diagnostic performance with a cut-off value of 174.5 kPa with the associated sensitivity and specificity of 100.0% and 80.3% respectively. Meanwhile, E-sd in system-2 displayed the best diagnostic performance with a cut-off value of 12.7 kPa, with the associated sensitivity and specificity of 94.4% and 80.3% respectively. The diagnostic performance of the two 2D-SWE systems was not statistically different according to receiver operating characteristic curve (ROC) analysis of E-max, E-mean, and E-sd. Conclusion: For identifying breast lesions, system-1 and system-2 appear to be similar in diagnostic performance. However, different cut-off values for different parameters might be selected to obtain the best diagnostic performance for the two 2D-SWE systems. |
Databáze: | MEDLINE |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |