Prospective evaluation of feasibility and electrophysiologic and echocardiographic characteristics of left bundle branch area pacing.

Autor: Vijayaraman P; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania. Electronic address: pvijayaraman1@geisinger.edu., Subzposh FA; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., Naperkowski A; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., Panikkath R; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., John K; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., Mascarenhas V; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., Bauch TD; Geisinger Heart Institute, Geisinger Commonwealth School of Medicine, Wilkes Barre, Pennsylvania., Huang W; Department of Cardiology, The First Affiliated Hospital, Wenzhou Medical University, Wenzhou, China.
Jazyk: angličtina
Zdroj: Heart rhythm [Heart Rhythm] 2019 Dec; Vol. 16 (12), pp. 1774-1782. Date of Electronic Publication: 2019 May 25.
DOI: 10.1016/j.hrthm.2019.05.011
Abstrakt: Background: His bundle pacing (HBP) is the most physiologic form of pacing but associated with higher thresholds and lower success in patients with His-Purkinje conduction disease. Recent reports have described transvenous left bundle branch area pacing (LBBAP).
Objective: We aimed to prospectively evaluate the feasibility and the electrophysiologic and echocardiographic characteristics of LBBAP.
Methods: Patients requiring pacing for bradycardia or heart failure indications (failed left ventricular [LV] lead) were prospectively enrolled. LBBAP was performed with a Medtronic 3830 lead. Presence of left bundle branch (LBB) potential, paced QRS morphology/duration, and peak LV activation time (pLVAT) were recorded at implant. Pacing threshold and sensing was assessed at implant and follow-up. Echocardiography was performed to assess the approximate lead location and impact on tricuspid valve function.
Results: LBBAP was successful in 93 of 100 (93%) patients. Mean age was 75 ± 13 years; men 69%, left bundle branch block 24%, right bundle branch block 25%, intraventricular conduction defect 8%. Indications for pacing were atrioventricular (AV) block 54%, sinus node dysfunction 23%, AV node ablation 7%, cardiac resynchronization therapy 11%, HBP lead failure 7%. Baseline QRS duration was 133 ± 35 ms. Paced QRS duration was 136 ± 17 ms. LBB potentials were observed in 63 patients with left bundle branch - ventricle (LBB-V) interval of 27 ± 6 ms. pLVAT was 75 ± 16 ms. Pacing threshold at implant was 0.6 ± 0.4 V @ 0.5 ms and R waves were 10 ± 6 mV and remained stable at median follow-up of 3 months. The lead depth in the septum was approximately 1.4 ± 0.23 cm.
Conclusions: LBBAP was feasible in a high percentage of patients with low thresholds during acute follow-up. HBP and LBBAP may significantly increase the overall success of physiologic pacing.
(Copyright © 2019 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.)