In-stent restenosis associated with dual-layer Roadsaver carotid artery stent: a retrospective single-center study.
Autor: | Ierardi AM; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Angileri SA; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Brambillasca PM; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Jannone ML; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Biondetti P; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Petrillo M; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy., Crippa M; Vascular Surgery Unit, Surgical Sciences Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Milan, Italy., Pinto A; Department of Radiology, CTO Hospital, Azienda dei Colli, Naples, Italy., Carrafiello G; Diagnostic and Interventional Radiology Department, ASST Santi Paolo e Carlo, San Paolo Hospital, University of Milan, Via A di Rudinì 8, 20142, Milan, Italy. gcarraf@gmail.com. |
---|---|
Jazyk: | angličtina |
Zdroj: | La Radiologia medica [Radiol Med] 2019 Jul; Vol. 124 (7), pp. 704-709. Date of Electronic Publication: 2019 Mar 02. |
DOI: | 10.1007/s11547-019-01019-7 |
Abstrakt: | Aim: To evaluate the incidence of in-stent restenosis (ISR) at 12-month follow-up, in patients treated with new dual-layer Roadsaver (Terumo Corp, Tokyo, Japan) carotid artery stent (CAS). Materials and Methods: Thirteen patients underwent CAS and received a Roadsaver. Neurological examination was performed in all patients. Carotid stenosis was revealed by Doppler ultrasound (DUS) and multidetector CT (MDCT) scan. Four patients presented a peak systolic velocity (PSV) between 130 and 150 cm/s, six a PSV between 150 and 180 cm/s, and three a PSV > 180 cm/s. MDCT gave further anatomic information. Direct stenting was performed in 11 (84.6%) cases, whereas in 2 (15.4%) cases predilatation was required. In all cases postdilatation was performed. Technical and clinical success and safety were evaluated. Stent patency was evaluated during the 12-month follow-up. Results: Technical success was achieved in all cases. In three patients a nonsignificant residual stenosis < 30% has been reported. No major complications during or after the procedure occurred. One patient (7.7%) showed a transitory bradycardia during angioplasty. One (7.7%) local bleeding at the puncture site treated conservatively occurred. All the neurological examinations performed 24 h and 30 days after the procedure were negative. No significant ISR was registered. In 2 patients, < 30% ISR was revealed at DUS performed after 6 months and confirmed at 12 months. In both patients CEUS and MDCT denied the presence of significant stenosis. Conclusions: Roadsaver stent seems to be durable. Further studies with longer-term outcome are necessary to confirm our results. |
Databáze: | MEDLINE |
Externí odkaz: |