[Reliability of an automatic monitor for blood pressure measurement].
Autor: | Cardona-Müller D; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Grover-Páez F; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Guzmán-Saldivar V; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Alanis-Sánchez GA; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Murguía-Soto C; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Totsuka-Sutto SE; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Quezada-Fernández P; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Macías-Chumacera A; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Cardona-Muñoz EG; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México., Ramos-Becerra CG; Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México. |
---|---|
Jazyk: | Spanish; Castilian |
Zdroj: | Revista medica de Chile [Rev Med Chil] 2018 Feb; Vol. 146 (2), pp. 190-195. |
DOI: | 10.4067/s0034-98872018000200190 |
Abstrakt: | Background A correct blood pressure (BP) measurement is essential for the diagnosis and control of high BP. Aim: To evaluate the agreement and repeatability of BP measurements with the OMRON HEM-7320-LA device compared to a mercury sphygmomanometer. Material and Methods: A cross-sectional study comparing BP measurements made by two randomly selected trained nurses and an automatic oscillometric device. The mercurial sphygmomanometer was connected to the automated device via a "T" type connector and a dual-head stethoscope was used, allowing simultaneous measurements. The results were analyzed with one-factor analysis of variance, Bland-Altman's test, repeatability coefficient (RC), and intra-class correlation coefficient (ICC). Results: Forty-nine participants aged 56 ± 19 years were included. Nineteen had hypertension (38%). We did not observe a significant difference in either systolic (SBP) or diastolic blood pressure (DBP) pressure measurements between the observers and the device. The mean difference was -0.09 mmHg (95% confidence intervals (CI)-0.9 to 0.7) for SBP and -0.9 mmHg (95% CI -1.7 to -0.13) for DBP. The RC for SBP (6.2, 5.2 and 5.8 mmHg) and DBP (4.7, 4.2 y 5.2 mmHg) was similar between the observers and the device. The ICC for SBP was 0.990 (95% CI 0.983 to 0.995, p < 0.01) and 0.986 (95% CI 0.977 to 0.991, p < 0.01) for DBP. Conclusions: There was a high level of agreement and similar measurement repeatability in the measurements performed by the automatic device and the mercurial sphygmomanometer. No differences in BP measurements were observed. |
Databáze: | MEDLINE |
Externí odkaz: |